High precision calibration source DIGISTANT® for voltage,current and thermocouples, for precise test and  calibration and for control of processes, basic accuracy 0.005 %, RS232 and IEEE488 interface

Product Inquiry

    Product comparison

      • Features

        • High precision current and voltage source ± 52 mA, ± 30 V 
          Option: ± 22 mA, ± 60 V
        • Precision simulation for all conventional thermocouple types (optional)
        • Error limit 0.003 % Rdg.
        • Standard with RS232 and IEEE488 interface, USB and Ethernet (optional)
        • Current "SINK"
      • Product data

        Model 4462
        DC voltage simulation

        300 mV
        3 V
        30 V
        60 V (Option)

        Measurement accuracy voltage simulation 0.003 %
        DC voltage measurement No
        DC current simulation

        52 mA
        (22 mA)

        Measurement accuracy current simulation 0.007 %
        DC current measurement No
        Thermocouples simulation Yes
        RTD simulation No
        Interfaces RS232
        IEEE488
        Supply voltage

        230 V / 45 - 65 Hz
        115 V / 45 - 65 Hz

        Specific characteristics

        • High precision current and voltage source ± 52 mA, ± 30 V 
        • Option: ± 22 mA, ± 60 V
        • Precision simulation for all conventional thermocouple types (optional)
        • Error limit 0.003 % Rdg.
        • Standard with RS232 and IEEE488 interface, USB and Ethernet (optional)
        • Current "SINK"
      • Description

        The precision calibration unit combines high accuracy, low drift, low noise and superior long-term stability with multiple functionality and simple operation. Ramps, Δ+/Δ-, and multiple setpoint storage make the operation of the device easier for the user.

        For that reason the application possibilities are many:

        • Calibration of current and DC-voltage meters
        • Precise testing of thermocouple temperature measuring instruments
        • Calibration of controllers, sensors, detection devices and other devices used in process control
        • Open-loop process control with the aid of integrated ramp function

        DIGISTANT® model 4462 can be used both as a standalone table-top device, as well as in automatic, computerassisted manufacturing and testing systems.

        It is possible to set currents of ± 200 nA ... ± 52 mA, voltages of ± 1 μV ... ± 30 V and, optionally temperature setpoint values of 14 thermocouple types.

        The output value is fed back via the sensor line to eliminate voltage drops across the measuring leads. The device has an adjustable current/voltage limitation. An external voltage divider of 1 up to 1:1000 can be considered internally.

        With the thermovoltage sourcing option you can enter °C, °F and K, the temperature scales ITS 90 or IPTS 68 and the comparison point mode constant/external. Furthermore, when sourcing thermocouples a calibrated external comparison point can be used, whereby the data for calibration in the device can be taken into consideration.

        Indication of the source value is carried out in large 12 mm figures on an illuminated graphics-LCD.

        The device can be operated both via the keyboard as well as the interface.

      • Data Sheet & Downloads

        Data sheet
        Operation manual
        Software
      • Accessories

        pdf 2329-Z004
        pdf 2329-Z006
        pdf External reference junction for thermocouples Model 4485-V001


      burster in USA

      burster Inc.
      Adam Giorgione, Christina Giorgione: +1-706-331-0537; Darren McCulley: +1-330-842-5088
      adam.giorgioneburster.us; christina.giorgioneburster.us; darren.mcculleyburster.us; salesburster.us
      www.burster.us

      Product category :

      • Torque Sensors
      • Load Cells
      • Displacement Sensors
      • Mechanical and Electrical Calibration Solutions

      Company-Information Company-Information

      Physicom Corp.
      +1 4167543168
      salesphysicomcorp.com
      www.physicomcorp.com

      Product category :

      • Press-Fit Control / DIGIFORCE®
      • Process Instruments

      Company-Information Company-Information

      USA

      burster Inc.

      Adam Giorgione; Christina Giorgione
      105 Laurel Creek Rd. SE
      Suite: 10C Calhoun, GA. 30701

      Darren McCulley
      304 State Route 303
      44241 Streetsboro OH

      USA

      Fon: Adam Giorgione, Christina Giorgione: +1-706-331-0537; Darren McCulley: +1-330-842-5088
      adam.giorgioneburster.us; christina.giorgioneburster.us; darren.mcculleyburster.us; salesburster.us
      www.burster.us

      Product category :

      • Torque Sensors
      • Load Cells
      • Displacement Sensors
      • Mechanical and Electrical Calibration Solutions

      Physicom Corp.

      P.O. Box 462
      Scarborough, ON
      M1E 4Y9
      Kanada

      Fon: +1 4167543168
      Fax: +1 4167542351
      salesphysicomcorp.com
      www.physicomcorp.com

      Local sales support through different rep networks available, please inquire at Physicom.

      SERVICES

      • Qualified design, application support
      • Process optimization & rental equipment
      • Qualified commissioning, integration support & training
      • On-site trouble shooting & update service
      • Qualified phone & online support
      • Full repair facilities
      • Local calibration service

      Product category :

      • Press-Fit Control / DIGIFORCE®
      • Process Instruments
      ×
      © 2024 // burster präzisionsmesstechnik gmbh & co kg
      nach oben